The device log file and the replaced circuit board therapy control unit m16.2 were available for investigation.Based on the log entries the reported event could be confirmed.A reboot was conducted due to a communication disruption between the two microprocessors on the m16.2.After the reset the device returned to the previous valid operating state.During a test run of the replaced circuit board no comparable reboot could be recorded in the logbook nor a hardware fault was detected during a functional test.Although no persistent hardware defect could be determined, the circuit board m16.2 could be identified as root cause for the reported reboot.A reset of the processors on the m16 therapy control unit leads to a interruption of ventilation for a maximum of 15 seconds.Subsequently ventilation will be continued with the last valid settings.If the reset time will be exceeded, an acoustic alarm is generated via the secondary alarm system.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|