AESCULAP AG CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.; LIGATION / VESSEL CLIPS Back to Search Results
Model Number PL574T
Device Problem Mechanical Jam (2983)
Patient Problem Insufficient Information (4580)
Event Date 02/20/2023
Event Type  malfunction  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with pl574t - challenger ti-p sm-ligat.Clips 12 cartr.According to the complaint description, the clip jammed during the surgery.The pusher bar come off.Patient harm was unknown.Additional information was not provided nor available.The malfunction is filed under aag reference (b)(4).
 
Manufacturer Narrative
Additional information: d9 -no product return h6 - codes updated investigation: the complaint sample was not provided for investigation, thus an investigation of the complaint product is not possible.Device history review: the device quality and manufacturing history records have been checked for all leading device(s) lot numbers and the products found to be according to our specification valid at the time of production.There are no similar complaints against the same lot number(s) with this error pattern.Conclusion/preventive measures: due to a lack of data and without the product we cannot determine an exact conclusion and root cause for the mentioned deviation.According to the quality standard, a production error and a material defect can most probably be excluded.If further investigations are required, the product should be provided for examination.Root cause cannot be finally concluded.Therefore, the root cause specific risk cannot be identified.The potential risk determined during initial vigilance evaluation remains valid.Based upon the investigation results, a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.
Type of Device
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der gruen
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key16436627
MDR Text Key310409311
Report Number9610612-2023-00043
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K081031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPL574T
Device Catalogue NumberPL574T
Device Lot Number52799618
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/20/2023
Initial Date FDA Received02/24/2023
Supplement Dates Manufacturer Received10/05/2023
Supplement Dates FDA Received10/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-