• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-55
Device Problem No Device Output (1435)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during testing the cs300 intra-aortic balloon pump (iabp) there is a printer malfunction.There was no patient involvement.
 
Event Description
It was reported that during testing the cs300 intra-aortic balloon pump (iabp) there is a printer malfunction.There was no patient involvement.However, there was no injury & adverse event reported.
 
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue.To fix the issue, the fse replaced the printer recorder and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released for cleared for clinical service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16445846
MDR Text Key310341442
Report Number2249723-2023-01294
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567112541
UDI-Public10607567112541
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-55
Device Catalogue Number0998-00-3023-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/12/2010
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-