Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/06/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source: australia.Product is in process of being returned to zimmer biomet for the investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during a hip procedure, the threads of the cup inserter snapped off of the inner shafter during impaction of the cup.There was approximately 15 mins of delay in the surgery with no harm or injury to the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Event Description
|
No additional event information to report at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional and/or corrected information.Visual examination of the provided pictures identified the threaded tip has fractured.Lot identification is necessary for review of device history records, lot identification was not provided.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|