• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH HI-TEC, INC. AIA-2000; FLUOROMETER, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH HI-TEC, INC. AIA-2000; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-2000 ST
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/31/2023
Event Type  malfunction  
Manufacturer Narrative
A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed the complaint by reviewing the error logs and reproduced error by running a test sample.Fse checked the clog sensitivity on the pressure board of the main arm and found the offset sensitivity value was lower than specification and attempted to adjust it but the value would not change.Fse restarted the analyzer, but the offset value still did not change and could not adjust it.Fse replaced the pressure board and still could not adjust the offset sensitivity.Fse then detached the tubing to the sample nozzle and was then able to adjust the sensitivity.Fse found the sample nozzle clogged and replaced the sample nozzle when cleaning didn't seem to help.Fse resolved the complaint by replacing the sample nozzle and adjusting the clog sensitivity fse validated the analyzer by successfully performing quality control run without error and within acceptable range.No further action required by field service.The aia-2000 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were three (3) similar complaints identified during the searched period, which includes this event.The aia-2000 operator's manual under appendix 4: error messages states the following: [2070] clogging detected during specimen suction by main arm cause : the negative pressure detected after specimen suction exceeded the standard.The specified amount of specimen may not have been obtained because the sampling nozzle was blocked.The measurement result will be flagged (sc flag).Solution : verify that the specimen is free of solid substances (such as fibrin) or that there is sufficient volume of specimen if it is prepared in the primary tube, and retry the measurement.If retry fails, contact tosoh service center or local representatives.The most probable cause of the reported event is due to clog in sample nozzle and faulty pressure board.
 
Event Description
A customer reported getting error message "2070 clogging detected during specimen suction by main arm" on the aia-2000 analyzer.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for estradiol (e2), follicle stimulating hormone (fsh) and progesterone (progiii).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key16458090
MDR Text Key310989838
Report Number3004529019-2023-00268
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189284616
UDI-Public04560189284616
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000 ST
Device Catalogue Number022100
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/31/2023
Initial Date FDA Received03/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-