• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 100 SC - DEU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 100 SC - DEU Back to Search Results
Model Number 27051
Device Problem Capacitative Coupling (1079)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2023
Event Type  malfunction  
Manufacturer Narrative
The astral device was returned to resmed.Evaluation confirmed the reported complaint.Visual inspection of the main circuit board revealed a leaky super capacitor.The main circuit board was replaced to address the reported complaint.The device was serviced and fully tested before it was returned to the customer.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed an error message (sf179) related to the super capacitor.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 100 SC - DEU
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key16461606
MDR Text Key310490087
Report Number3007573469-2023-00136
Device Sequence Number1
Product Code NOU
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27051
Device Catalogue Number27051
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/01/2023
Distributor Facility Aware Date01/30/2023
Device Age91 MO
Date Report to Manufacturer03/01/2023
Date Manufacturer Received01/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2015
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-