Catalog Number 391.771 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/08/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Initial reporter occupation: reporter is a j&j employee.Device evaluated by mfr: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from synthes reports an event in japan as follows: it was reported that this was a spinal fusion with the uss schanz for the vertebral body fracture on (b)(6) 2023.After final locking, when the surgeon cut the unnecessary part of screw with the bolt cutting, the bolt cutting was broken.The bolt cutting could not be used, so that the hospital¿s instruments were used for cutting.The surgery was completed successfully within 30 minutes delay.There no were adverse consequence/s that affected the patient because of the reported event.Patient status/outcome: is stable.No further information is available.Concomitant product details: unk - screws: trauma (part # unknown, lot # unknown, quantity involved - unknown).This report is for one (1) bolt cutting head ø5 long cutting-h 2 f/.This is report 1 of 1 for complaint (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that bolt cutting head ø5 long cutting-h 2 f/ was broken from the distal tip.The broken fragment was returned for examination.No other issues were found.A dimensional inspection for the bolt cutting head ø5 long cutting-h 2 f/ was not performed as is not applicable to the complaint condition.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the bolt cutting head ø5 long cutting-h 2 f/ would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: yes.Dimensional inspection: n/a.Device history lot =part # 391.771, synthes lot # 4551691, supplier lot # 1157772, release to warehouse date: 20 feb 2003, supplier: (b)(4).No ncrs were generated during production.Device history review: review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|