• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE SERFAS ENERGY 90-S CRUISE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE SERFAS ENERGY 90-S CRUISE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0279401200
Device Problem Continuous Firing (1123)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/06/2023
Event Type  malfunction  
Event Description
It was reported that the device continuously activated.
 
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: continuous activation probable root cause: probe short, button stuck on firing position, fluid ingress, suction tube cut, button inadvertently pressed, damaged insulation, probe bender used to bend nonbendable probe, user accidentally submerges device, inadequate gluing operations (tip and core/lumen) or inadequate gluing/welding of core to handle console error.Use error.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
 
Event Description
It was reported that the device continuously activated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERFAS ENERGY 90-S CRUISE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16461748
MDR Text Key310496603
Report Number0002936485-2023-00178
Device Sequence Number1
Product Code GEI
UDI-Device Identifier07613327057034
UDI-Public07613327057034
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K160050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 07/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0279401200
Device Catalogue Number0279401200
Device Lot Number22201AE2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-