|
Lot Number 0029661319 |
Device Problems
Use of Device Problem (1670); Air/Gas in Device (4062)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/02/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during a cryoablation procedure to treat atrial fibrillation (afib), a polarsheath was selected for use.After the sheath was inserted inside the patient, the catheter was inserted into the sheath and air was confirmed during a reverse blood flow confirmation.The physician slowly applied negative pressure, but it could not be resolved even when performed several times.There was no problem after the sheath was replaced.The sheath will be returned for analysis.
|
|
Manufacturer Narrative
|
This polarsheath was received by boston scientific and laboratory analysis was performed.Visual inspection did not reveal any abnormalities.Standard aspiration tests performed revealed all results were within specifications.Hemostasis and air pressure testing was conducted, and the sheath met all specifications with no signs of leaking or disruption of pressure.Laboratory analysis was unable to confirm the reported clinical observations; the polarsheath passed all relevant testing.The instructions for use state "do not aspirate via the side port if the sheath lumen is occupied (i.E., by the dilator or components of the cryoablation catheter) as the aspiration may draw air across the sheath valve into the polarsheath.
|
|
Event Description
|
It was reported that during a cryoablation procedure to treat atrial fibrillation (afib), a polarsheath was selected for use.After the sheath was inserted inside the patient, the catheter was inserted into the sheath and air was confirmed during a reverse blood flow confirmation.The physician slowly applied negative pressure, but it could not be resolved even when performed several times.There was no problem after the sheath was replaced.The device has been received for analysis.
|
|
Search Alerts/Recalls
|
|
|