• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 30°, 4 MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 30°, 4 MM Back to Search Results
Model Number A22002A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/16/2022
Event Type  malfunction  
Manufacturer Narrative
The suspect device was sent to an olympus service center for evaluation.Inspection and testing found the system tube sunk down.The reported issue (cracked lens) was confirmed during testing due to a broken lens in the optical system.In addition, there were dents on the distal frame due to collision with other devices, there were dents on the eyepiece funnel due to collision with other devices, the black laser marking was faded, the investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
The customer reported the lens of the subject device was found to be cracked during reprocessing.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information that was received about the event.Sections e2, e3, and g2 were updated.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the broken components could not be determined.It is possible that the event occurred due to the application of excessive force by the user.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 30°, 4 MM
Type of Device
TELESCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, . 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16463386
MDR Text Key310952904
Report Number9610773-2023-00621
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761020855
UDI-Public04042761020855
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K897003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22002A
Device Catalogue NumberA22002A
Device Lot Number684259
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/19/2022
Initial Date FDA Received03/01/2023
Supplement Dates Manufacturer Received03/23/2023
04/21/2023
Supplement Dates FDA Received04/20/2023
04/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-