Model Number A22002A |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/16/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device was sent to an olympus service center for evaluation.Inspection and testing found the system tube sunk down.The reported issue (cracked lens) was confirmed during testing due to a broken lens in the optical system.In addition, there were dents on the distal frame due to collision with other devices, there were dents on the eyepiece funnel due to collision with other devices, the black laser marking was faded, the investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|
|
Event Description
|
The customer reported the lens of the subject device was found to be cracked during reprocessing.There was no harm or user injury reported due to the event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information that was received about the event.Sections e2, e3, and g2 were updated.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the broken components could not be determined.It is possible that the event occurred due to the application of excessive force by the user.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|