Catalog Number 8606000 |
Device Problems
Gas Output Problem (1266); Failure to Deliver (2338); Device Sensing Problem (2917)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/09/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that a ventilator failure occurred during use.There was no detail in the report which would reasonably suggest that patient consequences may have occurred.
|
|
Manufacturer Narrative
|
The investigation has just started - results will be provided in a follow-up report.Investigation is on-going.
|
|
Manufacturer Narrative
|
The device was subject to an on-site evaluation by an engineer of the local dräger s&s organization.The reported vent fail during use could be confirmed upon checking the log file.The records stored therein indicate that the supervisor function of the device forced a shutdown of automatic ventilation due to a too high power consumption of the ventilator motor.It could be further determined that this was caused by piston movement against the limit stop because the detection system for the motor position had obviously a problem.After replacement of the connection cable for the light barrier - part of the position detection system - the device was fully functional again and could be returned to use.Due to the low field failure rate, an in-depth investigation of the cable was not considered necessary.Dräger finally concludes that the device responded as designed upon a certain error condition - an autonomous shut-down was forced to protect from serious damages to the ventilator and other components.The user was alerted to this condition by means of a corresponding alarm; manual ventilation including gas dosage remains possible.
|
|
Event Description
|
It was reported that a ventilator failure occurred during use.There was no detail in the report which would reasonably suggest that patient consequences may have occurred.
|
|
Search Alerts/Recalls
|