DEPUY ORTHOPAEDICS INC US PINN CAN BONE SCREW 6.5MMX15MM; BONE SCREWS AND PINS : SCREWS Back to Search Results
Model Number 1217-15-500
Device Problem Difficult to Advance (2920)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2023
Event Type  malfunction  
Event Description
It was reported that the 30mm screw head sheered off on insertion.The 15mm screw did not seat after the surgeon used a 30mm drill bit, a seated drill guide, and a 15mm screw.Surgeon commented that something has changed with our screws.30mm screw threads were left in the patient but the 30mm screw head was removed.It was unknown if there was a surgical delay.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device 121715500 lot d22052823 and no non-conformances / manufacturing irregularities were identified.
 
Event Description
Additional information was received and stated that there was a surgical delay of 3 minutes while attempting to remove the threads.Response to patient harm was unknown and broken threads were left in patient.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN CAN BONE SCREW 6.5MMX15MM
Type of Device
BONE SCREWS AND PINS : SCREWS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JJM (SUZHOU) LTD. 3006356043
no.299 changyang st
suzhou industrial park
suzhou jiangsu 21512 -6
CH   21512-6
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16467663
MDR Text Key310512600
Report Number1818910-2023-04891
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295009306
UDI-Public10603295009306
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K983014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 03/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1217-15-500
Device Catalogue Number121715500
Device Lot NumberD22052823
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age49 YR
Patient SexFemale
Patient Weight65 KG
-
-