Catalog Number 530.710 |
Device Problems
Break (1069); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Additional narrative: the actual device was returned for evaluation.The battery oscillator device was evaluated and the reported condition that the oscillating head was off axis was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that moving parts of the trigger of the device did not move smoothly, it was further determined that the device failed pretest for check for sticky triggers.The assignable root cause of this condition was determined to be traced to component failure due to wear.Udi: (b)(4).
|
|
Event Description
|
It was reported from germany that during service and evaluation, it was determined that the moving parts of the trigger of the battery oscillator device did not move smoothly, the saw head was loose, and the device had insufficient/low power.It was further determined that the device failed pretest for check positioning of the saw head, check for sticky trigger, and check oscillation frequency with frequency meter.It was noted in the service order that the oscillating head was off axis.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|