The investigation was made based on the log file.It revealed an interruption of the communication between the therapy control unit (m16.2) and the guic.As consequence a reboot was forced and a system failure alarm given.As identified by the service technician a loose contact within the cable connection (dvi cable) can be determined as root cause for the reported event.In case of a reset of the processors of the therapy control unit m16.2 therapy is discontinued for a maximum of 15 seconds, before therapy will be resumed with the latest valid settings.If for any reasons therapy was not restarted within 15 seconds an acoustical alarm will be given.Moreover, manual ventilation remains possible.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|