Model Number 27003 |
Device Problems
Inadequate User Interface (2958); No Tactile Prompts/Feedback (4024)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2023 |
Event Type
malfunction
|
Event Description
|
During resmed evaluation, an astral device had an unresponsive touchscreen.There was no patient harm or serious injury reported as a result of this incident.
|
|
Manufacturer Narrative
|
The top case was replaced to address the reported complaint.The device was serviced and fully tested before it was returned to the customer.Resmed reference#: (b)(4).
|
|
Manufacturer Narrative
|
Based on all available evidence and complaint investigations of a similar nature, an investigation determined that the reported complaint was due to internal physical damage.Resmed¿s risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
|
|
Event Description
|
During resmed evaluation, an astral device had an unresponsive touchscreen.There was no patient harm or serious injury reported as a result of this incident.
|
|