Model Number 27003
Device Problems Inadequate User Interface (2958); No Tactile Prompts/Feedback (4024)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2023
Event Type  malfunction  
Event Description
During resmed evaluation, an astral device had an unresponsive touchscreen.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
The top case was replaced to address the reported complaint.The device was serviced and fully tested before it was returned to the customer.Resmed reference#: (b)(4).
 
Manufacturer Narrative
Based on all available evidence and complaint investigations of a similar nature, an investigation determined that the reported complaint was due to internal physical damage.Resmed¿s risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Event Description
During resmed evaluation, an astral device had an unresponsive touchscreen.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - AMER
Type of Device
VENTILATOR, CONTINUOUS (FACILITY/HOME)
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key16478993
MDR Text Key310679524
Report Number3007573469-2023-00143
Device Sequence Number1
Product Code NOU
UDI-Device Identifier00619498270033
UDI-Public(01)00619498270033(11)181114(10)1343438
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27003
Device Catalogue Number27003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA07/18/2023
Distributor Facility Aware Date06/20/2023
Device Age50 MO
Date Report to Manufacturer07/18/2023
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/03/2023
Supplement Dates Manufacturer Received06/20/2023
Supplement Dates FDA Received07/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/19/2018
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-