Updated fields - b5, h4, h6.This report is being supplemented to correct and provide additional information as reflected on b5.The suspect device was returned to olympus for evaluation and the allegation was confirmed.When the slider was pressed, the needle would not stick out and the tube was buckling.There were no other abnormalities noted that led to the event reported.The device history record for the subject device was reviewed.No abnormalities were noted which could cause or contribute to the reported problem.The suspect device was returned to olympus for evaluation and the allegation was confirmed.When the slider was pressed, the needle would not stick out and the tube was buckling.There were no other abnormalities noted that led to the event reported.A definitive root cause could not be determined.However, based on the results of the device evaluation and the available information, the investigation determined the following likely occurred: 1.) friction resistance increased between the outer tube and the needle tube due to buckling of the tube, causing the reported event.2.) a bending force was possibly applied to the tube when the device was inserted into the endoscope, removed from the sterile package or during pre-inspection, causing the tube to buckle.As a preventive measure, the instructions for use provides the following statements: ¿the operator of this instrument must be a physician or medical personnel under the supervision of a physician and must have received sufficient training in clinical endoscopic technique.This manual does not explain or discuss clinical endoscopic procedures.¿before use, prepare and inspect the instrument as instructed below.Should any irregularity be observed, do not use the instrument; use a spare the instead.Damage or irregularity may compromise patient or user safety; for example, posing an infection control risk, causing tissue irritation, perforation, bleeding, or mucous membrane damage, and may result in more severe equipment damage.¿do not coil the insertion portion with a diameter of less than 15 cm.This could damage the insertion portion.¿before use, confirm that the needle and the insertion portion are not damaged.If any irregularity such as significant deformations or excessive bends is found, do not use the instrument.Otherwise, it may cause perforation, bleeding, mucous membrane damage.¿do not advance or extend the instrument abruptly.This could cause patient injury, such as perforation, bleeding, or mucous membrane damage.It could also damage the endoscope and/or instrument.¿when inserting the instrument into the endoscope, retract the needle into the tube, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.¿stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.
|