• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDHFA500Q
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Heart Failure/Congestive Heart Failure (4446)
Event Date 11/01/2022
Event Type  Death  
Event Description
Related manufacturing reference number: 2017865-2023-11749.Related manufacturing reference number: 2017865-2023-11750.Related manufacturing reference number: 2017865-2023-11752.It was reported that the patient passed due to congestive heart failure and hypertensive arteriosclerotic heart disease.There is no known allegation from a healthcare professional that the biventricular implantable cardioverter defibrillator system caused or contributed to the patient's death.
 
Manufacturer Narrative
The device was returned due to patient death.Final analysis found it was above elective replacement indicator (eri) when received.The device was found to communicate appropriately with a merlin programmer and has not reached the elective replacement indicator (eri).Ate testing was performed and the device was normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16483232
MDR Text Key310673081
Report Number2017865-2023-11746
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Model NumberCDHFA500Q
Device Catalogue NumberCDHFA500Q
Device Lot NumberP000145506
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/14/2023
Initial Date FDA Received03/03/2023
Supplement Dates Manufacturer Received05/18/2023
Supplement Dates FDA Received05/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age65 YR
Patient SexMale
Patient Weight93 KG
-
-