• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: A.I.D.D LONGFORD MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

A.I.D.D LONGFORD MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Model Number 08P1934
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2023
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event Description
The customer observed a falsely decreased alinity c magnesium result for one patient.The result was not reported out.The sample was repeated and the result was higher.The following data was provided processed on 14feb2023: sid (b)(6) initial result = 0.48 repeat result = 1.12.Reference range = 0.66 to 1.07 mmol/l no negative impact to patient management was reported.
 
Manufacturer Narrative
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Return testing was not completed as returns were not available.A search for similar complaints and ticket trending review did not identify any trends or issues with the complaint lot number or current complaint issue.Device history record review did not identify any non-conformances or deviations related to the complaint lot and complaint issue.The historical performance of reagent lot 43489ud00 was evaluated using data from customers worldwide.The patient data was analyzed and compared to an established control limit.This evaluation indicated that the patient median result for lot 43489ud00 is within the established control limits.Therefore, no unusual reagent lot performance was identified for lot 43489ud00.Labeling was reviewed and was found to adequately address the issue under review.Based on the investigation, no systemic issue or deficiency was identified for the alinity c magnesium reagent, lot 43489ud00.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Type of Device
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key16483458
MDR Text Key311481873
Report Number3005094123-2023-00061
Device Sequence Number1
Product Code JGJ
UDI-Device Identifier00380740176532
UDI-Public00380740176532
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K181748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/17/2023
Device Model Number08P1934
Device Catalogue Number08P19-34
Device Lot Number43489UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/15/2023
Initial Date FDA Received03/04/2023
Supplement Dates Manufacturer Received04/07/2023
Supplement Dates FDA Received04/12/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/12/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALNTY C PROCESSING MODU, 03R67-01, AC02346; ALNTY C PROCESSING MODU, 03R67-01, AC02346
-
-