Catalog Number 254183000 |
Device Problems
Material Deformation (2976); Unintended Movement (3026); Illegible Information (4050)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/17/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that while attempting to make the mold, it was found that the three pieces of the mold would not fit together.All pieces are being returned.No surgical delay.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Additional information received states that it appeared that the holes may have been burred up preventing the peg from going into it.
|
|
Manufacturer Narrative
|
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # : (b)(4).Investigation summary : the device associated with this report was returned to depuy synthes for evaluation.Visual examination of the device confirmed the complaint.Visual examination found scratches and nicks throughout of the mold, this condition prevents the etch from being visible.Additionally, burr and wear were observed at the assembly orifices, this condition causes the parts of the mold to not have a tight fit.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a manufacturing records evaluation (mre) was not performed as no lot number was able to be retrieved for this device.
|
|