• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO CUSTOM TUBING PACK; TUBING, PUMP, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO CUSTOM TUBING PACK; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HY1W12R36
Device Problem Fluid/Blood Leak (1250)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 02/10/2023
Event Type  Injury  
Event Description
Medtronic received information that during use of a custom tubing pack, it was reported that the vrv valve was leaking. there was no patient impact associated with this event.Additional information received stated that the patient lost about 200ml of blood.
 
Manufacturer Narrative
Device evaluation summary: upon receipt in our quality analysis lab, visual inspection showed no outward signs of any damage.The device had a test mark, which confirmed the vrv was tested prior to distribution.Functional testing of the returned device was performed at 0.5 l/min.The instructions for use provides an approximation for first time use of negative and positive pressures to open the umbrella relief valves at 0.5 l/min blood flow are approximately -205 mmhg and 318 mmhg.Our results indicated the negative pressure umbrella opened at -180 mmhg and the positive pressure umbrella valve opened at 364 mmhg verifying the functionality for the valves for the returned device.The acceptable function of the umbrella valves also confirms the duck bill valve functioned as intended.There were no leaks observed during operation.Reason for return was not confirmed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information b5:medtronic received additional information that the same leak did not appear in multiple packs.There was no visible air in system/tubing, there was blood from the suction lines coming into the reservoir that would seep out from the top of the reservoir.The customer gave three units of blood post-op to the patient.The additional information for b5 that is included above was received on the 07 march 2023.The information in the conclusion below is from the 01 may 2023.Conclusion: the complaint is unconfirmed for leaking valve.There was no observed damage to the device and performance testing indicated that the device functions as intended.After evaluation it was determined there was no product failure that occurred during the reported event.The device history record was not reviewed as the returned product analysis found no evidence of manufacturing issues with the returned device.Complaints received for the same failure mode were reviewed and showed no trends warranting escalation related to this occurrence.Assessment against the medtronic risk management file process failure mode effects and criticality analysis (pfmeca) document indicated that the current risk zone does not exceed the risk zone predicted in the product pfmeca.Medtronic will continue to monitor for future occurrences and trends.Correction b5: there was no adverse patient effect associated with this event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM TUBING PACK
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16494090
MDR Text Key310782484
Report Number9612164-2023-01015
Device Sequence Number1
Product Code DWE
UDI-Device Identifier00763000535841
UDI-Public00763000535841
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K171979
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHY1W12R36
Device Catalogue NumberHY1W12R36
Device Lot Number225398679
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-