U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS ANTI-HBC II; TEST, HEPATITIS B

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS ANTI-HBC II; TEST, HEPATITIS B Back to Search Results
Catalog Number 07374160190
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/28/2023
Event Type  malfunction  
Manufacturer Narrative
Calibration and controls were acceptable.No reagent issue is observed.Upon review of the alarm trace, no relevant alarms were observed.The patient sample material was requested, but could not be provided.Based on product labeling, samples with a cutoff index > 1.0 are non-reactive with the elecsys anti hbc ii assay.These samples are considered negative for anti-hbc and do not need further testing.The investigation could not identify a product problem.The cause of the event could not be determined.Na.
 
Event Description
The initial reporter stated they received questionable non-reactive results for one patient sample tested with the elecsys anti-hbc gen.2 immunoassay on a cobas 8000 e 602 module (serial number (b)(4).No questionable results were reported outside of the laboratory.A previous sample of the patient collected on (b)(6) 2023, resulted in a "nucleic acid test (nat) - hiv,hcv, hbv" result of non-reactive when tested using the cobas s201 multiplex pcr method.The complained sample was initially tested twice on the e 602 analyzer with anti-hbc reagent lot 600333 on (b)(6) 2023, resulting in anti-hbc values of 2.23 coi (non-reactive) and 2.24 coi (non-reactive).The complained sample was tested using the abbott alinity ci-series anti-hbc method on 28-jan-2023, resulting in a value of 1.21 s/co (reactive).The complained sample was re-centrifuged and repeated twice on the e 602 analyzer with anti-hbc reagent lot 600333 on (b)(6) 2023, resulting in anti-hbc values of 2.23 coi (non-reactive) and 2.27 coi (non-reactive).The complained sample was tested using the vitros anti-hbc method on (b)(6) 2023, resulting in a value of 3.65 (negative).No units of measure were provided for this result.The complained sample was tested on the e 602 analyzer with anti-hbc reagent lot 633393 on (b)(6) 2023, resulting in an anti-hbc value of 2.31 coi (non-reactive).
 
Manufacturer Narrative
The component code has been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS ANTI-HBC II
Type of Device
TEST, HEPATITIS B
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16496382
MDR Text Key310882165
Report Number1823260-2023-00702
Device Sequence Number1
Product Code LOM
Combination Product (y/n)N
Reporter Country CodeSA
PMA/PMN Number
P100031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07374160190
Device Lot Number600333, 633393
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/06/2023
Initial Date FDA Received03/07/2023
Supplement Dates Manufacturer Received03/08/2023
Supplement Dates FDA Received03/31/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age39 YR
Patient SexMale
-
-