Other, other text: h6: health impact, event methods, evaluation, and conclusion codes: updated.One warmer device was received for investigation.During visual inspection no damage or anomalies were observed with the device.The device over-temperature alarm was functionally tested, and the device alarm activated within specification.A review of job records related to the reported serial number indicated the device passed all functional and alarm testing and calibration.Based on the available information and the results of the testing, the investigation could not confirm the reported complaint or attribute a root cause.Preventative maintenance was performed on the device.
|