Model Number PL574T
Device Problem Unintended Ejection (1234)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2023
Event Type  Injury  
Manufacturer Narrative
Investigation on-going.Additional information/investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with pl574t - challenger ti-p sm-ligat.Clips 12 cartr.According to the complaint description, the clip jammed on the second shot during the surgery.The cartridge came off and was partially chipped, but was recovered.An additional medical intervention was necessary.Additional information was not available.The adverse event is filed under aag reference (b)(4).Involved components: pl574t - challenger ti-p sm-ligat.Clips 12 cartr.- lot 52775168 pl574t - challenger ti-p sm-ligat.Clips 12 cartr.- lot 52775168.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.
Type of Device
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key16504225
MDR Text Key310898054
Report Number9610612-2023-00050
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K081031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPL574T
Device Catalogue NumberPL574T
Device Lot Number52778601
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PL574T - LOT 52775168.; PL574T - LOT 52775168.
Patient Outcome(s) Required Intervention;
-
-