|
Model Number EMAX2PLUS |
Device Problems
Material Frayed (1262); Unintended System Motion (1430)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition that the device had unintended activation/motion, identified during service and repair, was confirmed.The assignable root cause was determined to be due to component failure from wear.Udi: (b)(4).
|
|
Event Description
|
It was reported by india that during service and evaluation, it was determined that the motor device had unintended activation/motion, cord damage and produced e5(handpiece fault) error code.It was further determined that the device failed pretest for visual assessment, cable assessment and safety assessment.It was noted in the service order that during pre-surgery, it was discovered that the device had a hole in the hose.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|