Model Number CI-1400-01 |
Device Problems
Electrical /Electronic Property Problem (1198); Electro-Static Discharge (2149); Intermittent Communication Failure (4038)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 03/16/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Event Description
|
The recipient is reportedly experiencing a loos of lock.External equipment was exchanged, however the issue did not resolve.Device testing could not be completed due to loss of lock.Revision surgery will be scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed sliced silicone overmold on the top and bottom covers, as well as a severed electrode.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock could not be obtained at any spacing.The electrode condition prevented an electrical test from being performed.The device passed the mechanical tests performed.The failure of this device is attributed to a short from a power node to the case ground at the analog chip.It is believed that electrostatic discharge (esd) led to an overload voltage, damaging structures on case ground node inside the analog integrated circuit.This ultimately caused the device to cease functioning.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|