• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO ENDO DISSECT; MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO ENDO DISSECT; MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED Back to Search Results
Model Number 176645
Device Problem Failure to Align (2522)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
According to the reporter, the teeth did not line up.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO DISSECT
Type of Device
MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16513590
MDR Text Key311061474
Report Number2647580-2023-00925
Device Sequence Number1
Product Code GET
UDI-Device Identifier10884523000740
UDI-Public10884523000740
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K904578
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number176645
Device Catalogue Number176645
Device Lot NumberP2L0440
Was Device Available for Evaluation? No
Date Manufacturer Received02/22/2023
Date Device Manufactured11/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-