It was reported that, during the implant of this inflatable penile prosthesis (ipp), the device was successfully inflated with a surrogate reservoir, but then could not be deflated.Multiple attempts of deflation were attempted but were unsuccessful.The fully inflated device was explanted and replaced with another ipp.No patient complications were reported.
|
Upon receipt of the pump and cylinders of this inflatable penile prosthesis (ipp) at our quality assurance laboratory, the returned components underwent a thorough analysis.The cylinders were visually inspected and functionally tested.No damage was identified and both cylinders passed leak testing.The pump was visually inspected and functionally tested.No leaks were found.The pump was functionally tested and did not pass deflation testing.Product analysis confirmed a functional issue with the pump.Based on the information available and analysis results, analysis confirmed the reported clinical observation of a deflation issue.
|