Model Number AVSM10060 |
Device Problems
Loss of or Failure to Bond (1068); Positioning Failure (1158)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/20/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a stent graft placement procedure for a 67 year old patient with a basilic swing point stenosis in an av fistula, the stent allegedly did not deploy at all.It was further reported that the stent was removed.The procedure was completed using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
A voluntary recall has been initiated for the covera vascular covered stent which was product catalog/lot number specific.Reportedly the covera vascular covered stent has the potential to exhibit deployment issues (i.E., failure to deploy the covered stent) due to slide block bond failures in the device handle.Full or partial failure of the product to deploy is most likely to require percutaneous replacement with existing access.In rare cases, iatrogenic vascular injury may occur as the result of device withdrawal, manipulation, or abrupt/unexpected movements, particularly in the context of difficult or partial stent deployment or mispositioning.Adjunctive endovascular maneuvers may be indicated.There have been no cases where an open surgical procedure has been required.To date, there have been no reported patient injuries associated with deployment issues.As a result of the field action, this event is being reported as a malfunction reportable event.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiration date: 10/2023).
|
|
Event Description
|
It was reported that during a stent graft placement procedure for a 67 year old patient with a basilic swing point stenosis in an av fistula, the stent allegedly did not deploy at all.It was further reported that the stent was removed.The procedure was completed using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
A voluntary recall has been initiated for the covera vascular covered stent which was product catalog/lot number specific.Reportedly the covera vascular covered stent has the potential to exhibit deployment issues (i.E., failure to deploy the covered stent) due to slide block bond failures in the device handle.Full or partial failure of the product to deploy is most likely to require percutaneous replacement with existing access.In rare cases, iatrogenic vascular injury may occur as the result of device withdrawal, manipulation, or abrupt/unexpected movements, particularly in the context of difficult or partial stent deployment or mispositioning.Adjunctive endovascular maneuvers may be indicated.There have been no cases where an open surgical procedure has been required.To date, there have been no reported patient injuries associated with deployment issues.As a result of the field action, this event is being reported as a malfunction reportable event.Manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent delivery system of a covered stent was returned for evaluation.The slide block, a force transmitting component was no longer connected to the proximal sheath.The disconnection of the slide block is considered to led to the reported impossibility to deploy the stent which leads to confirmed results.It was reported that an 8fr introducer and a 0.035" guidewire was used during the procedure.Based on evaluation of the sample the inability of an adhesive joint (slide block/tether/diving sheath) to withstand tension force during deployment is confirmed.A definitive root cause could not be determined based upon the available information.Labeling review: the relevant labeling for this product was reviewed.Based on the instructions for use supplied with this product delivery system specific events that could be associated with clinical complications include but are not limited to failure to deploy and high deployment forces.With regards to preparation and accessories, the instructions for use states "pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated".Holding and handling of the system throughout deployment was found sufficiently described.H10: d4 (expiration date: 10/2023), g3, h6 (device).Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|