Catalog Number ABC4466BA
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone number (additional): (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the individual packaging of thirty-one (31) minicaps were damaged; further described as ¿minicaps came out soft¿ and ¿they come almost uncovered." this was observed prior to use of the devices for peritoneal dialysis therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the actual devices were not available; however, fourteen (14) retention samples were evaluated at the plant.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.The reported condition was not verified on the retention samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CUERNAVACA
ave. de los 50 metros no. 2
cuernavaca 62578
MX   62578
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16521407
MDR Text Key311090361
Report Number1416980-2023-00968
Device Sequence Number1
Product Code KDJ
UDI-Device Identifier00085412353036
UDI-Public(01)00085412353036
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/31/2024
Device Catalogue NumberABC4466BA
Device Lot NumberM22I28A
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/13/2023
Initial Date FDA Received03/10/2023
Supplement Dates Manufacturer Received04/17/2023
Supplement Dates FDA Received04/18/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-