The customer reported to olympus, after reprocessing with oer-4 reprocessor, the bronchovideoscope had a green foreign substance attached to the distal part of the scope and connector.When other scopes were reprocessed, there was no adherence of foreign matter.So the customer determined that the issue was with the scope.The cleaning, disinfection and sterilization (cds) process followed onsite included primary cleaning followed by high level disinfection with oer-4 reprocessor.The issue occurred twice with the same scope.There was no patient involvement.This report captures complaint on the first occurrence.Patient identifier (b)(6) captures complaint on second occurrence of the same event.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause cannot be specified.The event can be detected/prevented by following the instructions for use which state: 3 inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.10 if foreign objects, such as detergent remnants, hard water residue, finger grease, dust, and lint may be on the electrical contacts on the endoscope connector (ex.Wiping with lint-prone cloths, left unused for a long period of time), wipe the electrical contacts with clean lint-free cloths moistened with 70% ethyl or 70% isopropyl alcohol.Also, confirm that the electrical contacts are completely dry and clean.Olympus will continue to monitor field performance for this device.
|