H4 manufacturing date ¿ added.H3 device evaluated by mfg ¿updated.H3 summary attached - updated.D4 expiration date - added.D9 product available to stryker- updated.D9 returned to manufacturer on- updated.Due to the automated manufacturing execution system (mes) system there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection, the stent was seen to be deployed.The stent delivery wire (sdw) was seen to be kinked.The stent struts were seen to be broken and the stent was seen to be deformed.The introducer sheath was returned intact.The functional inspection could not be performed as the stent was returned deployed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported event of stent broken/fractured during use was confirmed during analysis.The reported event of stent difficult/unable to advance or pullback through catheter and stent difficult/unable to transfer could not be replicated during device analysis; however, the analysis results are consistent with the reported event.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.Additional information received was the device was prepared as per the directions for use, there was no damage noted to the packaging prior to opening the packaging, the device was confirmed to be in good condition prior to use on the patient and continuous flush was set up and maintained throughout the clinical procedure.The device was analyzed.The stent was returned in the condition of deployed, however it was reported the stent was deployed on the table by the physician.The stent was also found to be deformed and broken/fractured.The sdw was returned and found to be kinked and the introducer sheath was returned and was found to be intact.Its likely when the reported resistance was felt while attempting to transfer the stent during the procedure manipulation of the device, would have caused the damage noted to the device during analysis.The reported events stent difficult/unable to advance or pullback through catheter, stent difficult/unable to transfer and stent broken/fractured during use as well as the analyzed defects stent broken/fractured during use, sdw kinked/bent and stent deformed will be assigned procedural factors as this complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.
|