• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE; RIGID OPTICAL LAPAROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE; RIGID OPTICAL LAPAROSCOPE Back to Search Results
Model Number WA53005A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned and evaluated.The evaluation confirmed the damaged eyepiece.Dents were also noted on the outer tube.A review of the device history record (dhr) found no deviations that could have caused or contributed to the reported issue.The dhr confirmed that the subject device was shipped in accordance with the specifications.A definitive root cause for this issue was not established, however, it is probable that the issue occurred because the device was subjected to excessive stress forces/mishandled by the end-user.Olympus will continue to monitor field performance for this device.
 
Event Description
The customer reported to olympus that the rigid endoscope had a cracked ocular/eyepiece.There were no reports of patient or user harm associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE
Type of Device
RIGID OPTICAL LAPAROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16525354
MDR Text Key311508226
Report Number9610773-2023-00725
Device Sequence Number1
Product Code EWY
UDI-Device Identifier04042761052764
UDI-Public04042761052764
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K912362
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA53005A
Device Catalogue NumberWA53005A
Device Lot Number656497
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2023
Was the Report Sent to FDA? No
Date Manufacturer Received02/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-