Catalog Number 6903190
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/13/2023
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.Device not returned.
 
Event Description
It was reported that during a dialysis catheter placement procedure, the catheter allegedly had a laceration in one of the catheter pathways and air was allegedly noticed in the catheter.It was further reported that blood had allegedly came out at the site of the catheter that was found lacerated.There was no reported patient injury.
 
Event Description
It was reported that during a dialysis catheter placement procedure, the catheter allegedly had a laceration in one of the catheter pathways and air was allegedly noticed in the catheter.It was further reported that blood had allegedly came out at the site of the catheter that was found lacerated.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.Four electronic photos were provided for review.The photos show one equistream split tip catheter implanted in a patient body.Therefore, the investigation is inconclusive for the reported material puncture and leak issue as there was no objective evidence obtained from the provided photos.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 12/2023).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUISTREAM SPLIT TIP
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16527230
MDR Text Key311255613
Report Number3006260740-2023-00774
Device Sequence Number1
Product Code MSD
UDI-Device Identifier00801741121647
UDI-Public(01)00801741121647
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K090101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 05/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6903190
Device Lot NumberREGQ1178
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/13/2023
Initial Date FDA Received03/12/2023
Supplement Dates Manufacturer Received05/22/2023
Supplement Dates FDA Received05/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/19/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-