Model Number 441916 |
Device Problem
False Negative Result (1225)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that bd max¿ system, bd max¿ instrument experiencing negative samples and when run on geneexpert positive results.The following information was provided by the initial reporter: this lab has started to validate the bd max respiratory assay they are running same samples in geneexpert what they can see is that some samples are negative in bd max and positive in geneexpert.
|
|
Manufacturer Narrative
|
Pma/510k info: k111860, k130470.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
|
Manufacturer Narrative
|
After further evaluation of the complaint, it has been determined that the previously submitted report 1119779-2023-00272 was sent in error.There was no report of serious injury, medical intervention, or reportable device malfunction.Therefore this is not considered to be a reportable malfunction.
|
|
Event Description
|
It was reported that bd max¿ system, bd max¿ instrument experiencing negative samples and when run on geneexpert positive results.The following information was provided by the initial reporter: this lab has started to validate the bd max respiratory assay.They are running same samples in geneexpert.What they can see is that some samples are negative in bd max and positive in geneexpert.
|
|