• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. WEB SL; INTRASACCULAR FLOW DISRUPTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. WEB SL; INTRASACCULAR FLOW DISRUPTION DEVICE Back to Search Results
Model Number UNKNOWN
Device Problems Retraction Problem (1536); Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Manufacturer Narrative
The lot number was not provided; therefore, a search for associated production-related non-conformances could not be performed.The device was implanted and not available for return and investigation.Images were provided.However, the investigation is ongoing.Upon completion of the investigation, a supplemental mdr will be submitted.
 
Event Description
It was reported that during a procedure for a large ruptured basilar tip aneurysm, the web failed to re-sheath or re-open when the physician attempted to adjust.This resulted in a mushroom head shaped web inside of the aneurysm dome and a long stem down the basilar artery.The physician collapsed the web device into the aneurysm and used a y stent and coils to successfully complete the surgery.There was no reported patient injury. .
 
Manufacturer Narrative
Six radiographic images are supplied in place of a returned product.The images show the distal web is expanded in the distal part of the aneurysm.The proximal part (not detached) is collapsed and protrudes into the basilar tip.The images do not identify the cause for this event.Without the return and physical evaluation of the device, the investigation cannot determine if a condition exists that would have caused on contributed to the reported event.Items returned: n/a.Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination could not be performed as this information was not available at the time this investigation was performed.Ifu review (additional information can be found in the ifu): potential complications potential complications include but are not limited to the following: vessel puncture site hematoma, aneurysm perforation or rupture, hemorrhage, edema, thromboemboli, transient ischemic attack, ischemic stroke, neurologic deficits, parent artery occlusion, ischemia, vessel dissection or perforation, vascular thrombosis, vasospasm, device migration or misplacement, premature detachment, headache, post-embolization syndrome, infection and death.The web embolization device requires the use of fluoroscopy.Potential complications related to angiographic and fluoroscopic radiation doses include, but are not limited to, alopecia, burns ranging in severity from skin reddening to ulcers, cataracts, and delayed neoplasia.The probability of occurrence of complications may increase as procedure time and number of procedures increase.Other procedural complications including but not limited to anesthetic and contrast media risks, hypotension, hypertension and access site complications.Warnings: do not rotate the delivery device during or after delivery of the web embolization device.Rotating the web embolization device may result in damage or premature detachment.Procedure - instructions for use detachment of the web embolization device 34.The detachment control device is pre-loaded with batteries and will activate when the delivery device is properly connected.35.Verify that the rhv is firmly locked around the delivery device before attaching the detachment control device to ensure that the web embolization device does not move during the connection process.36.Ensure that the delivery device gold connectors are clean and free from blood or contrast.If necessary, wipe the connectors with sterile water and dry before connecting.37.Insert the proximal end of the delivery device into the detachment control device.When the delivery device is properly connected, the light will flash green and an intermittent tone will be heard.38.Verify the web embolization device position before pressing the detachment button.39.Push the detachment button.During firing, the light should be solid green and the beep should be continuous.40.Verify detachment by first loosening the rhv valve, then pulling back slowly on the delivery device and verifying that there is not web embolization device movement.If the web embolization device does not detach, push the detachment button again.If the web embolization device is still not detached, obtain a new detachment control device and attempt detachment up to two additional times.If it does not detach, remove the delivery device.41.Verify the position of the web embolization device angiographically through the guide catheter.42.Prior to removing the microcatheter from the treatment site, place an appropriately sized guidewire completely through the microcatheter lumen to ensure that no part of the web embolization device remains within the microcatheter.The physician has the discretion to modify the web embolization device deployment technique based on the complexity and variation in embolization procedures.Any modifications must be consistent with the previously described procedures, warnings, precautions and patient safety information in these instructions for use.
 
Event Description
Please see h10 for investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEB SL
Type of Device
INTRASACCULAR FLOW DISRUPTION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key16534979
MDR Text Key311218954
Report Number2032493-2023-00613
Device Sequence Number1
Product Code OPR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P170032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/20/2023
Initial Date FDA Received03/13/2023
Supplement Dates Manufacturer Received03/16/2023
Supplement Dates FDA Received03/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-