• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON® SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON® SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 55840007550
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having deformity correction.The pre-op diagnosis was ais.It was reported that, during post-op the distal implant(screw) was sliding from end of rod.Screw was dislocated.Patient underwent revision, the screw was explanted and a new screw and set screw were implanted.One month later the new screw that was replaced was also found sliding out from the rod on (b)(6) 2023.There is a plan to perform another revision surgery for this patient.There were no patient symptoms reported as a result.No further complications reported regarding the event.
 
Manufacturer Narrative
Date of event was unknown.Country of event: south korea.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Section e: facility name and address has been corrected.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Radiographic image review comments: 1.Lateral x-ray for long segmental tl posterior fusion.Overall contour appears quite flat.Rod out of one on interior screws.2.Ap image of some on one side.Thoracic and lumbar screws/rod connected with domino.3.Multiple ap images dated "noted as (b)(6) 2022 demonstrate possible worsening of curvature.Rod on one side appears to lengthy through last screw." 4.Ap image of some, no date, rod barely passed through last screw on one side.5.Rod present through bottom screw level.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON® SPINAL SYSTEM
Type of Device
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key16535520
MDR Text Key311214424
Report Number1030489-2023-00167
Device Sequence Number1
Product Code KWP
UDI-Device Identifier00643169004351
UDI-Public00643169004351
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K113174
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number55840007550
Device Catalogue Number55840007550
Device Lot NumberH5683690
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/21/2023
Initial Date FDA Received03/14/2023
Supplement Dates Manufacturer Received02/21/2023
03/15/2023
Supplement Dates FDA Received03/21/2023
04/04/2023
Date Device Manufactured05/13/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age20 YR
Patient SexMale
-
-