Model Number AUTOLUBE-III |
Device Problems
Break (1069); Material Fragmentation (1261)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that the foot control device had broken gauge cover.During in-house engineering evaluation, it was observed that the device had a ruptured hose, damaged gauge, and component damage.It was further determined that the device was unable to control irrigation flow rate.It was further determined that the device failed pretest for visual assessment, gauge condition and drip rate assessment.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the incident was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition of the broken gauge cover was not confirmed.Therefore, an assignable root cause was not determined.However, the ruptured hose, identified during service and evaluation was confirmed.The assignable root cause was determined to be traced to user, which is user error.Udi: (b)(4).
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition of the broken gauge cover was confirmed.The assignable root cause was determined to be traced to user, which is user error.
|
|
Search Alerts/Recalls
|
|
|