PERFUSION SYSTEMS OXY/CVR COMBO BB841 FUSION W/BALANCE 26L; RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number BB841 |
Device Problems
Obstruction of Flow (2423); Air/Gas in Device (4062)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/14/2023 |
Event Type
malfunction
|
Event Description
|
Medtronic received information that during use of a fusion cardiotomy venous reservoir, it was reported that the de-foamer overflowed and/or was obstructed from the recirculation line coming in.The device was used to complete the procedure.There was no reported adverse patient effect associated with this event.Medtronic received additional information that there was a possible obstruction inside the reservoir.Medtronic received additional information that the device was used to complete the procedure, it was moved to another port.
|
|
Manufacturer Narrative
|
Device evaluation: visual inspection showed no outward signs of any damage/assembly errors.During the cleaning process there were no flow issues observed.The reason for return was undetermined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information b5: medtronic received additional information that foaming subsided once foam hit the top of the cardiotomy venous reservoir and the venous air from the cannulation site subsided.Air was observed in the venous line.Vacuum assisted venous drainage (vavd) was used.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|