Additional information was added to d9, h3, h4, h6 and h10.H10: the device was received for evaluation.A visual inspection was done which revealed fluid inside the housing which indicated a leak occurred.Further examination revealed the cause of the leak was due to missing film-wrap.The film-wrap is located at the upper area of the bladder and fastens the bladder to the stress member post.When the film-wrap is missing, the bladder will not inflate during fill and fluid will go directly inside the housing.The reported condition was verified.The cause of the condition was due to an assembly issue during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|