• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25RWC W/ RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25RWC W/ RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX25RWC
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2023
Event Type  malfunction  
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, a leakage occurred at the point of the arterial thermistor.*no health consequences or impact.*product was changed out.*procedure completed successfully.
 
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.Adverse event problem:component code: 4739 - gas exchanger; health effect - impact code: 2645 - no patient involvement; health effect - clinical code: 4582 - no clinical signs, symptoms or conditions; medical device problem code: 1354 - leak/splash; investigation findings: 3233 - results pending completion of investigation.Investigation conclusions: 11 - conclusion not yet available.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information and device evaluation).H3 (device evaluated by manufacturer).H4 (device manufacture date).H6 (identification of evaluation codes 10, 11, 3331, 4210, 25).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #2: 3331 - analysis of production records.Investigation findings: 4210 - leakage/seal.Investigation conclusions: 25 - cause traced to manufacturing.The returned sample was inspected upon receipt, no visual anomalies were noted.The unit was pressurized with water and was found to leak immediately from the arterial thermistor.A representative retention sample was previously used for testing and no anomalies were noted with the device prior to and during testing.A training was conducted with the production associates to make them aware of this event and to ensure that all thermistors, and other components, are properly assembled and inspected on the reservoir and oxygenator.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on march 15, 2023.Upon further investigation of the reported event, the following information is new and/or changed: h3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX25RWC W/ RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key16544529
MDR Text Key311747680
Report Number1124841-2023-00069
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450837
UDI-Public(01)00699753450837
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 04/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3CX*FX25RWC
Device Catalogue NumberN/A
Device Lot Number2K15
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received04/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-