ACCURAY INCORPORATED CYBERKNIFE TREATMENT DELIVERY SYSTEM; MEDICAL LINEAR ACCELERATOR Back to Search Results
Model Number G4
Device Problem Unintended System Motion (1430)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2023
Event Type  malfunction  
Manufacturer Narrative
Customer alleged the standard treatment couch has excessive movement when rolling left to right.No patient was involved, and no injury was reported.Investigation is ongoing.
 
Event Description
Standard treatment couch (stc) had excessive movement when rolling left to right.
 
Manufacturer Narrative
It has been determined that this issue occurred due to field service engineer error in performing the stc (standard treatment couch) set screw investigation as part of the pm that occurred 78 days prior to this issue being reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYBERKNIFE TREATMENT DELIVERY SYSTEM
Type of Device
MEDICAL LINEAR ACCELERATOR
Manufacturer (Section D)
ACCURAY INCORPORATED
1209 deming way
madison WI 53717
Manufacturer (Section G)
ACCURAY INCORPORATED
1209 deming way
madison WI 53717
Manufacturer Contact
daniel biank
1209 deming way
madison, WI 53717
6088242971
MDR Report Key16550413
MDR Text Key311379860
Report Number3003873069-2023-00003
Device Sequence Number1
Product Code IYE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K170788
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial,Followup
Report Date 01/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG4
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-