Model Number G4 |
Device Problem
Unintended System Motion (1430)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/16/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Customer alleged the standard treatment couch has excessive movement when rolling left to right.No patient was involved, and no injury was reported.Investigation is ongoing.
|
|
Event Description
|
Standard treatment couch (stc) had excessive movement when rolling left to right.
|
|
Manufacturer Narrative
|
It has been determined that this issue occurred due to field service engineer error in performing the stc (standard treatment couch) set screw investigation as part of the pm that occurred 78 days prior to this issue being reported.
|
|