Date of event is estimated.Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined. .
|
Related manufacturer reference number: 1627487-2023-01171 and 1627487-2023-01172.It was reported that the patient experienced ineffective therapy and pain at the ipg site.Troubleshooting revealed high impedance on both extension due to migration of the ipg.As a result, surgical intervention was undertaken on (b)(6) 2023 wherein the ipg was repositioned, and the extensions were explanted and replaced to address the issue.
|