• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JUNI OX FB FEM SZ 6 RM LL; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JUNI OX FB FEM SZ 6 RM LL; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71422356
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thromboembolism (2654)
Event Date 02/17/2023
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, after a ukr surgery was performed on (b)(6) 2022, the patient returned from an overseas flight and felt presyncopal and with pain in the right calf.The patient was later diagnosed with bilateral pulmonary emboli and extensive right lower leg dvts.The patient is recovering.
 
Manufacturer Narrative
H3, h6.The devices were not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, the provided electronic clinical report form documented the serious adverse event as ongoing and moderately severe which led to in-patient/prolonged hospitalization with admit/start date (b)(6) 2023 to discharge on (b)(6) 2023.The adverse event was categorized as an anticipated adverse event which was unrelated to the device, possibly related to the procedure, treated with medication therapies, and currently recovering/resolving.The electronic clinical report form concomitant medication/therapies shows ongoing therapies as of (b)(6) 2023 of enoxaparin (started on (b)(6) 2023) to treat and prevent blood clots as well as rivaroxaban to treat deep vein thrombosis and pulmonary emboli.A prior week of clexane was noted in the electronic clinical report form from 2nd to 9th of feb-2023 for deep vein thrombosis with ongoing ¿blood thinners¿ initiated on 9-feb-2023 that the patient does not remember the (medication) name.The patient is to follow up in clinic in approximately 6 months.Based on the information provided in the electronic clinical report forms, no further clinical factors could be definitively concluded to have contributed to the reported event; however, the ¿immobility from the long flight¿ could not be ruled out.Patient impact beyond the reported bilateral pulmonary emboli and extensive right lower leg deep vein thrombosis and subsequent medication along with hospitalization would not be anticipated as the event was reportedly resolved.No further medical assessment can be rendered at this time.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the tibial base and insert part numbers over the past 12 months and for the batch numbers based on historical data of the device did not reveal similar events for the listed device.A review of complaint history revealed similar events for the fixed bearing femoral over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed that thromboembolic diseases, including venous thrombosis, have been identified in the possible adverse effects.Besides, the warnings and precautions section revealed that postoperative patient care and directions and warnings to patients by physicians are extremely important.Protected weight bearing with external support is recommended for a period of time to allow healing.Normal daily activity may be resumed at the physician¿s direction.Patients should be directed to seek medical opinion before entering potentially adverse environments that could affect the performance of the implant.A review of the risk management files revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include post-operative healing issue and/or patient condition.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUNI OX FB FEM SZ 6 RM LL
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16558671
MDR Text Key311473473
Report Number1020279-2023-00566
Device Sequence Number1
Product Code HSX
UDI-Device Identifier03596010627421
UDI-Public03596010627421
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
K073175
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71422356
Device Catalogue Number71422356
Device Lot Number19EM09387
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/17/2023
Initial Date FDA Received03/16/2023
Supplement Dates Manufacturer Received04/04/2023
Supplement Dates FDA Received04/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/10/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Treatment
71422255 / 21KM03032; 71422434 / 21KM09946
Patient Outcome(s) Required Intervention;
Patient Age67 YR
Patient SexMale
Patient Weight91 KG
-
-