For the investigation the provided logfile was analysed.The log entries indicate that the vacuum pressure, that is required to keep the ventilator diaphragm in place, was too low.The software monitors several parameters that are relevant for a safe automatic ventilation, amongst others the membrane pressure (internal diaphragm vacuum).The fabius reacted as specified for this situation - stopped the automatic ventilation and posted a corresponding ventilator fail alarm.In this case, fresh gas delivery (incl.Agent) remains available and the patient can be ventilated using the manual breathing bag.There are several plausible root causes for such an issue with the vacuum pressure.On site, the dispatched service technician was able to reproduce the too low negative pressure by turning the a-cone switch during ventilation.However, the exact root cause could not be finally determined because also after calibrating the vacuum pump, the issue with the vacuum pressure was not solved.Thus, an intermittent leakage at a hose, e.G.At the a-cone switch or at the apl bypass valve, or a problem with the vacuum pump itself could be the root cause.Therefore, it has been recommended to check the hoses and the a-cone switch again and, if the problem cannot be solved, to replace the vacuum pump.
|