• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problems No Display/Image (1183); Fluid/Blood Leak (1250)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported that the warmer was leaking and broken display.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Manufacturer Narrative
Date of event is unknown, no information has been provided to date.A product sample was received and is awaiting evaluation, investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Other, other text: h3.Device evaluated by manufacturer and h6.Health impact, event problem and evaluation codes: updated.One (1) device was received.Visual inspection noted a broken liquid crystal display (lcd), cracked tank cover, and corroded drain fitting.Filled tank with water, attached temp check, plugged in line cord, and turned on the power switch.The liquid crystal display (lcd) was damaged, and the tank cover was leaking confirming the customer complaint.A root cause was due to impact on the liquid crystal display (lcd) and cracks in the tank cover.A manufacturing device history record (dhr) review was not performed because there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.Action taken, replaced the printed circuit board (pcb) which contains the liquid crystal display (lcd) and the tank cover.Performed preventative maintenance (pm).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16567936
MDR Text Key311569331
Report Number3012307300-2023-02679
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier10695085002796
UDI-Public10695085002796
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberHL-390
Was Device Available for Evaluation? No
Date Returned to Manufacturer03/01/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/27/2023
Initial Date FDA Received03/17/2023
Supplement Dates Manufacturer Received03/21/2023
Supplement Dates FDA Received03/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/17/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-