It was reported that the left ventricular (lv) lead pacing impedance (pli) measurement of this cardiac resynchronization therapy pacemaker (crt-p) system was greater than 3000 ohms, which was out-of-range (oor).This triggered a lead safety switch (lss) and a switch to unipolar sensing configuration.Technical services (ts) analyzed presenting electrogram (egm).There was a successful lv automatic threshold test confirming no loss of capture.Further evaluation was recommended.Patient was tested in clinic including isometric testing and the oor impedance could not be reproduced.This device was reprogrammed.It was also noted that this patient experienced muscle stimulation which was resolved with reprogramming.No additional adverse patient effects were reported.Currently, this crt-p system remains in service.
|