MAUDE Adverse Event Report: INSTITUT STRAUMANN AG PALATAL IMPLANT Ø4.1, L 4.2MM; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSTITUT STRAUMANN AG PALATAL IMPLANT Ø4.1, L 4.2MM; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Model Number Palatal-System
Device Problem Failure to Osseointegrate (1863)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2023
Event Type  Injury  
Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.  the non-integration of an endosseous dental implant during the healing phase is a known inherent risk of the treatment with dental implants.Most implant failures occur before occlusal loading (analainen et al.2009).Based on clinical studies about 1-3% of the implants fail within the first year after implantation (ganeles et al.2008).Implants may have to be removed in case one or more of the implant success criteria are not met.Implant success criteria according to buser et al.(1991) are: 1.Absence of persistent subjective complaints such as pain, foreign body sensation and /or dysesthesia.2.Absence of a recurrent peri-implant infection with suppuration.3.Absence of implant mobility.4.Absence of a continuous radiolucency around the implant.The manufacturer's trend analysis confirms that the reported early failure rate associated with its dental implants is below the expected failure rate for this treatment as published in the scientific literature.
 
Event Description
The clinician reports the implant was inserted (b)(6) 2022 in the patient's mouth.On (b)(6) 2023, non-osseointegration was verified.The device was forwarded to the manufacturer.There were no reported patient injuries or complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALATAL IMPLANT Ø4.1, L 4.2MM
Type of Device
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ  04002
Manufacturer (Section G)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ   04002
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
9787472509
MDR Report Key16578321
MDR Text Key311650735
Report Number0009613348-2023-026018
Device Sequence Number1
Product Code OAT
UDI-Device Identifier07630031718167
UDI-Public07630031718167
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K060062
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 03/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberPalatal-System
Device Catalogue Number042.335S
Device Lot NumberEEG56
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date03/21/2023
Event Location Other
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age11 YR
Patient SexMale
-
-