• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ACCENT MRI DR RF; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ACCENT MRI DR RF; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2224
Device Problems Incorrect Measurement (1383); Compatibility Problem (2960)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/02/2023
Event Type  Injury  
Event Description
It was reported that the patient presented in clinic for follow up.Upon review it was noted that the pacemaker exhibited incorrect measurements and compatibility problem.Programming changes were performed.There were no patient consequences.
 
Manufacturer Narrative
Further information was requested but not received.
 
Event Description
New information received indicated that the lead was explanted.
 
Manufacturer Narrative
The reported event of premature discharge of battery was not confirmed.The device was above elective replacement indicator (eri) level when received.A longevity assessment was performed and found the device was in the normal range of operation with appropriate remaining longevity.Based on this information, there was no evidence of premature depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCENT MRI DR RF
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16584898
MDR Text Key311766298
Report Number2017865-2023-14319
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2015
Device Model NumberPM2224
Device Lot Number4633640
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/27/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-