Model Number 24658
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2023
Event Type  malfunction  
Manufacturer Narrative
Initial reporter facility name: (b)(6) hospital.Device evaluated by mfr: the complaint device was received for product analysis.The device was returned with the balloon protector in place on the device.For investigation purposes the investigator fully removed the protector without issue.A visual examination identified that the balloon was not subjected to positive pressure.The balloon material and blades of the device were visually examined, and no issues were noted.All blades were fully bonded to the balloon material.A visual and tactile examination found the shaft polymer extrusion to be completely detached at the guidewire port.This type of damage is consistent with excessive tensile force being applied to the device.A visual investigation identified no damage or any issues with the markerbands of the device.This concludes the product analysis.
 
Event Description
Reportable based on device analysis completed on 28feb2023.It was reported that the device had crossing difficulties.The 100% stenosed target lesion was located in the severely tortuous and severely calcified left superficial femoral artery.A 4.00mmx1.5cmx140cm small peripheral cutting balloon was selected for use.During the procedure, it was noted that the device was unable to cross the lesion.The procedure was completed with another of the same device.No complications were reported and the patient was in good condition after the procedure.However, device analysis revealed shaft polymer extrusion completely detached at the guidewire port.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL PERIPHERAL CUTTING BALLOON
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16585330
MDR Text Key311756977
Report Number2124215-2023-10543
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 03/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/04/2023
Device Model Number24658
Device Catalogue Number24658
Device Lot Number0026299690
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-