OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE; HIGH DEFINITION AUTOCLAVABLE TELESCOPE
|
Back to Search Results |
|
Model Number WA53000A |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to the olympus for evaluation and the customer¿s reported problem was confirmed, the light guide post was detached from the scope.The investigation is still in progress; however, if additional information becomes available, this report will be supplemented accordingly.In general, the end-user is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.Additional 501(k): k923982/k950076.
|
|
Event Description
|
The customer reported to olympus the quick lock, high definition autoclavable telescope has a broken component defect, ¿detachment of light guide connection the customer reported problem occurred when washing the telescope as the light guide connection was turned with force.The procedure was completed successfully without delay.No death or injury and no impact to patient or other has been reported to olympus.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the broken light guild connection can be attributed to user error, improper handling and application of excessive force.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|